AGF System

AGF 2.0: Flüssigkeitsvernebler mit Zweistoffdüse und Zyklon (dp_max = 2 µm) nach VDI 3491-1 und -2. Generiert konstante Partikelrate und definiertes Partikelspektrum.

Liquid nebulizer with binary nozzle and downstream cyclone to separate large droplets

Liquid nebulizer with binary nozzle and cyclone (cut-off: 10 µm) as per VDI 3491-6
Liquid nebulizer with binary nozzle and cyclone (dp_max = 2 µm) as per VDI 3491-1 and -2
Aerosol generator with refill unit for loading compressed air filters under positive pressure up to 7 barg

The AGF series features a robust design, easy handling, and a long lifetime.

The AGF series combines a binary nozzle with a downstream cyclone. The AGF 2.0 is equipped with a cyclone with a cut-off diameter of dCut-off = 2 µm; the AGF 10.0 is equipped with a cyclone with a cut-off diameter of dCut-off = 10 µm.

The maximum particle concentration of the AGF series is approx. 107 particles/cm3. The cyclone can calculate the max. particle size, regardless of the liquid.

Large droplets generate high mass flows and may contaminate filters, for example, to such an extent that droplet loading causes the filters to be evaluated as better than they are during filter testing.

The AGF series is successfully used to generate test aerosols, in clean room technology applications, to test compressed air filters, and for LDV measurements.

Übersicht AGF System

Maße
BxHxT mm
Gewicht
Kg
Volumen
l/m
mmax* g/h dpmean*** µm dmax µm 115/230 V
50/60 Hz
Druckdicht bis zu 10 bar Druckluftzufuhr
AGF 2.0 300 x 325 x 175 Ca. 9 6 - 17 4 0,25 2 x
AGF 2.0 iP 300 x 325 x 175 Ca. 15 12 - 14 2 0,25 2 x
AGF 10.0 Ø240 x 385 Ca. 4 14 - 35 20 0,5 10 x
AGF 2.0 D Ø200 x 260 Ca. 8 12 - 45 4 0,25 2 x x
AGF 10.0 D Ø200 x 300 Ca. 8 14 - 35 20 0,5 10 x x
UGF 2000 270 x 200 x 175 Ca. 4 Ca. 1 - 13 1,5 0,2 1,5 x

Table 1: Overview of the AGF and UGF systems *applied for DEHS **test rig version ***average number diameter