PMFT 1000 M
The PMFT 1000 M tests face masks better than the standards EN 149/EN 13274-7 with additional accurate analysis of filter mask efficiency for SARS-CoV-2 (size approximately 120 nm to 160 nm). Both total photometric penetration and fractional efficiency are tested, e.g., the efficiency in the whole size range, respectively, the particle size-dependent penetration.
Norms and certificates
National and international standards
CCF (Covid Certified Filter) EN 149 EN 13274-7 GB 2626 42 CFR 84
Operation principle
PMFT 1000 M for reliable continuous operation in routine control for half masks
The measurement of total penetration and penetration via particle size is carried out with the high-precision aerosol photometer Promo® LED 2300.
The size distribution of the test aerosol according to the standard is as follows:
EN 149
Oil: Media diameter 290 nm | Geom. standard deviation 1.85
Salt: Media diameter 70 nm | Geom. standard deviation 2.5
42 CFR 84 / GB 2626
Oil: Media diameter 200 nm | Geom. standard deviation 1.6
Salt: Media diameter 55 nm | Geom. standard deviation 1.86
Based on the measurement of the particle penetration via the particle size, the PMFT 1000 M can also be used to consider the tolerances in the particle size distribution according to EN 13274-7 as min./max. penetration value. In addition, comparative values to other manufacturers are possible (option).
Compared with other standards (42CFR84/NIOSH) is also easy with the 42CFR84 upgrade KIT.
Overview of standards for face mask penetration testing
EN 149 | EN 13274-7 | EN 13274-7 | GB 2626 | GB 2626 | 42CFR 84 | 42 CFR 84 | |
---|---|---|---|---|---|---|---|
Aerosol | see EN 13274-7 | NaCl | PaO | NaCl | PaO/DOP | NaCl | DOP |
Mean diameter | see EN 13274-7 | 0.06 – 0.1 µm | 0.29 – 0.45 µm | 0.055 – 0.095 µm | 0.165 – 0.205 µm | 0.055 – 0.095 µm | 0.165 – 0.205 µm |
Standard deviation | see EN 13274-7 | 2 – 3 | 1.6 – 2.2 | < 1.86 (by additional software module) | < 1.6 (by additional software module) | < 1.86 (by additional software module) | < 1.6 (by additional software module) |
Concentration | see EN 13274-7 | 4 –12 mg/m³ | 15 – 25 mg/m³ | < 200 mg/m³ | (50 mg/m³) < 200 mg/m³ | < 200 mg/m³ | < 200 mg/m³ |
Discharge | - | - | - | required | required | required | required |
Air flow | see EN 13274-7 | 95 l/min | 95 l/min | 85 ± 4 l/min | 85 ± 4 l/min | 85 ± 4 l/min | 85 ± 4 l/min |
Temperature | see EN 13274-7 | 22 ± 3 °C | - | 25 ± 5 °C | 25 ± 5 °C | 25 ± 5 °C | 25 ± 5 °C |
Rel. humidity | see EN 13274-7 | < 40 % | - | 20 – 40 % (by compressed air) | - | 20 – 40 % (by compressed air) | 20 – 40 % (by compressed air) |
Measurement device | see EN 13274-7 | Sodium flame photometer | Light scattering photometer | particle detector | particle detector | Light scattering photometer | Light scattering photometer |
Measuring time | see EN 13274-7 | 30 s | 30 s | lowest eff. during loading | lowest eff. during loading | lowest eff. during loading | lowest eff. during loading |
Pause time | see EN 13274-7 | 180 s | 180 s | lowest eff. during loading | lowest eff. during loading | lowest eff. during loading | lowest eff. during loading |
Exposition | 120 mg | 120 mg | 120 mg | 200 ± 5 mg | 200 ± 5 mg | 200 ± 5 mg | 200 ± 5 mg |
PMFT remarks | O.K. | O.K. | O.K. | O.K. with upgrade KIT | O.K. with upgrade KIT | O.K. with upgrade KIT | O.K. with upgrade KIT |
Extensions/Accessories
The PMFT 1000 M is equipped with aerosol generators for measuring penetration with oil and salt. Measurement procedures for quick quality control (short test) or testing according to standard (exposure test) are supplied. Operation and automatic printout of the measurement results are therefore easy, even for inexperienced users.Benefits
Your special advantages
- Test rig working principle better than EN 149 and EN 13274-7
- Equivalent to GB 2626, 42 CFR 84 and ASTM 2299-3 by additional software option
- Test of community masks equivalent to CWA 17553
- Meets the requirements for respirators specified by the CCF (Covid Certified Filter) quality seal
- Includes two aerosol generators for NaCl and for oil
- Testing of fractional efficiency, e.g., efficiency in whole size range of 145 nm up to 40 µm
- Exact analysis of filter and filter mask efficiency for SARS-CoV-2 (size approx. 120 nm up to 160 nm). Efficiency also displayed at 145 nm
- Future proof: Works with any kind of aerosol without adjustments
- Further measurement of differential pressure, e.g., as well within different face velocities to simulate test of breathing resistance
- Face velocity adjustable between 1.5 – 70 cm/s
- Product capable of fast quality assurance and continuous optimization in R&D (display of size distribution)
- Individual face mask adapter for your product
- Attractive two years maintenance package for availability of test rig
- Can be operated as Mas Q-Check with optional Mas-Q-Head
The software extension additionally offers:
- Display of penetration results of the entire tolerance range of the size distribution according to EN 13274-7
- Allows the comparison of different test institutes and test systems
- Facilitates certification
- Shows wide range of standards. Depending on the test operation – i.e., particle size distribution of the test aerosol – one and the same mask can perform very well or fail the test
Applications
Individual solutions for various industries
- Reliable continuous operation in routine monitoring of half masks
- Test of total penetration for respiratory masks
- Exact analysis of filter mask efficiency e.g., Coronavirus
- Can be operated as Mas-Q-Check with optional Mas-Q-Head
Technical features
In detail
- Aerosols
- Salts (e.g. KCl, NaCl), liquid aerosols (e.g. DEHS), latex particles (PSL)
- Test area of the medium
- 100 cm2
- Measuring range (total penetration)
- 0.0005 - 100 %
- Measurement range (size)
- 0.145 – 40 µm
- Volume flow
- 1 – 27 m3/h - pressurized operation
- Power supply
- 115 – 230 V, 50/60 Hz
- Installation conditions
- +10 – +40 °C
- Differential pressure measurement
- 0 – 1,200 Pa
- Inflow velocity
- 1.5 – 70 cm/s (others on request)
- Compressed air supply
- 6 – 8 bar
- Dilution factor
- 1 : 27
- Test conditions according to standard
- +19 – +23 °C
- Dimensions
- Approx. 1,800 • 600 • 900 mm (H • W • D)