GFG 1000
The aerosol generator for the generation of test aerosols similar to diesel soot
Technical parameters
(Please see Brochure GFG for detailed information)
- Particle material: carbon; others on request
- Carrier gas: argon; others on request
- Volume flow: 4 - 6 l/min
Primary pressure: 4 bar
Dilution gas: particle-free and dry compressed-air - Volume flow: 0 - 40 l/min
Primary pressure: 4 - 8 bar - Number concentration: > 107 P/cm3
- Mass flow: ca. 0.06 - 6 mg/h
- Particle size: primary 3 - 5 nm
- agglomerate approx. 0 - 150 nm - Operating time: several hours nonstop
- Dimensions (WxHxD): 470 x 125 x 435 mm
- Weight: 23 kg
- Power supply: 115/230 V; 50/60 Hz
With the spark generator GFG 1000 test aerosols similar to diesel soot can be generated from pure graphite with an adjustable particle size distribution of approx. 20 nm - 150 nm and low mass flows up to approx. 6 mg/h with a high reproducibility.
Application examples
- Determination of separation efficiency of media for diesel soot filters
- Inhalation experiments
- Toxicology
Particular advantages
- Very good short-term and long-term constancy of particle size and concentration
- Fast adjustable particle size distribution
- Particle structure similar to diesel soot
- Easy handling
- Reliable function
- Best reproducibility
- Low maintenance
- Reduces your operating expenses
Please see Brochure GFG for detailed information.